Mito Red Light MitoPRO 300X
Mito Red Light MitoPRO 300X: declared 130 mW/cm² (solar meter) · reported 55.9 (LLIA, 15.24 cm)
gap 57.0%Irradiance declarations and reported measurements stay attached to instrument, distance, statistic, geometry, source, and accuracy class. Default order follows disclosure completeness.
Numbers at different distances remain separate. An unstated statistic is labeled. A missing distance stays unqualified.
Same geometry only
Mito Red Light MitoPRO 300X: declared 130 mW/cm² (solar meter) · reported 55.9 (LLIA, 15.24 cm)
gap 57.0%Mito Red Light MitoADAPT MIN 4.0: declared 125 mW/cm² (solar meter) · reported 56.4 (LLIA, 15.24 cm)
gap 54.9%Mito Red Light MitoPRO 1500X: declared 160 mW/cm² (solar meter) · reported 72.5 (LLIA, 15.24 cm)
gap 54.7%Mito Red Light MitoPRO 1500+: declared 68 mW/cm² (spectroradiometer) · reported 68.2 (LLIA, 15.24 cm)
gap -0.3%Mito Red Light MitoPRO 300+: declared 60 mW/cm² (spectroradiometer) · reported 60.1 (LLIA, 15.24 cm)
gap -0.2%full five-field disclosure
9 devices · 11.1% with distance · 0% with a third-party number · 0% with a K-number
full five-field disclosure
8 devices · 0% with distance · 0% with a third-party number · 12.5% with a K-number
full five-field disclosure
12 devices · 33.3% with distance · 0% with a third-party number · 0% with a K-number
full five-field disclosure
19 devices · 5.3% with distance · 21.1% with a third-party number · 63.2% with a K-number
full five-field disclosure
28 devices · 82.1% with distance · 39.3% with a third-party number · 0% with a K-number
| Source | Instrument | Geometry |
|---|---|---|
| Ben Rawlings | Jeti Specbos 2501 | wearing plane |
| ITL | Ocean Optics QE65000 | 2m integrating sphere 4pi |
| Light Therapy Insiders | OHSP-350-IRF | surface wearing geometry |
| LLIA | BTS2048-VL-TEC | 7x7 grid 49 points |
| LLIA | spectroradiometer | peak in device footprint |