Disagreement record
Mito Red Light MitoPRO 300X
Two or more irradiance numbers carry incompatible method fields. No cross-distance normalization is applied.
Comparability status
comparable: gap 57.0% · comparability rating 3 on a 1–3 scale
averaginginstrumentdeclared vs measured
Evidence
Number passports
130 mW/cm2 C2 · declared instrument and distance
- Claim ID
- rli:mito-red-light:mitopro-300x:irradiance-declared:mitoredlight-com-92e549:r1
- Technique
- solar_meter · 15.24 cm · peak
- Band
- red nir combined
- Geometry
- peak in device footprint
“>130 mW/cm2” source quotation
38.2 mW/cm2 L · signed report lineage
- Claim ID
- rli:mito-red-light:mitopro-300x:irradiance-report-average:mitoredlight-com-8c5984:r1
- Technique
- spectroradiometer · 15.24 cm · average
- Band
- red nir combined
- Geometry
- area below leds
“Average Irradiance @ 6in | 38.2 mW/cm2” source quotation
55.9 mW/cm2 L · signed report lineage
- Claim ID
- rli:mito-red-light:mitopro-300x:irradiance-report-peak:mitoredlight-com-8c5984:r1
- Technique
- spectroradiometer · 15.24 cm · peak
- Band
- red nir combined
- Geometry
- peak in device footprint
“Peak Irradiance @ 6in | 55.9 mW/cm2” source quotation