Mito Red Light MitoADAPT MIN 4.0
Two or more irradiance numbers carry incompatible method fields. No cross-distance normalization is applied.
Comparability status
comparable: gap 54.9% · comparability rating 3 on a 1–3 scale
averaginginstrumentdeclared vs measured
Evidence
Number passports
125 mW/cm2 C2 · declared instrument and distance
- Claim ID
- rli:mito-red-light:mitoadapt-min-4-0:irradiance-declared:mitoredlight-com-8c5984:r1
- Technique
- solar_meter · 15.24 cm · peak
- Band
- red nir combined
- Geometry
- peak in device footprint
“>125 mW/cm2” source quotation
40.7 mW/cm2 L · signed report lineage
- Claim ID
- rli:mito-red-light:mitoadapt-min-4-0:irradiance-report-average:mitoredlight-com-8c5984:r1
- Technique
- spectroradiometer · 15.24 cm · average
- Band
- red nir combined
- Geometry
- area below leds
“Average Irradiance @ 6in | 40.7 mW/cm2” source quotation
56.4 mW/cm2 L · signed report lineage
- Claim ID
- rli:mito-red-light:mitoadapt-min-4-0:irradiance-report-peak:mitoredlight-com-8c5984:r1
- Technique
- spectroradiometer · 15.24 cm · peak
- Band
- red nir combined
- Geometry
- peak in device footprint
“Peak Irradiance @ 6in | 56.4 mW/cm2” source quotation